Novus SMARTMETER Uživatelský manuál

Procházejte online nebo si stáhněte Uživatelský manuál pro Senzory Novus SMARTMETER. NOVUS SMARTMETER User Manual Uživatelská příručka

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 4
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 0
NOVUS PRODUTOS ELETRÔNICOS LTDA. – WEBSITE: WWW.NOVUSAUTOMATION.COM - TEL: +55-51-3323-3600 1/4
SMARTMETER
PORTABLE TEMPERATURE METER - INSTRUCTIONS MANUAL – V1.0x
1 INTRODUCTION
SmartMeter is a portable temperature meter with high-contrast LCD.
The Smart Meter housing enables industrial environment or bench-
top operation and provides a belt carrying fixture. It supports 2
temperature sensors with independent and concurrent display of
measured values. It provides the capability of reading 0-50 mV
electric signals with configurable indication scale.
Difference between channels, relative measure, minimum, maximum
and average values, hold and alarm functions are also available. It
also features an auto power off resource in case the device is idle, in
order to extend battery life.
The device is supplied with a K-type thermocouple with 1.5 m cable.
1.1 UNPACKING
The meter is supplied in proper packaging that offers suitable
protection during transit and storage. Unpack with care. Upon
delivery, make sure that there is no damage to the instrument, and
that there are no missing items.
Within packaging, you should find, besides the manual, the meter and
a K-type thermocouple sensor.
2 SPECIFICATIONS
Input types and max. indication range
J thermocouple..........................................................- 100 to 760 ºC
K thermocouple.......................................................- 150 to 1.370 ºC
T thermocouple..........................................................- 160 to 400 ºC
E thermocouple.............................................................-90 to 720 ºC
N thermocouple ......................................................- 270 to 1,300 ºC
R thermocouple ........................................................- 50 to 1,760 ºC
S thermocouple.........................................................- 50 to 1,760 ºC
B thermocouple.........................................................500 to 1,820 ºC
Pt100 .........................................................................- 200 to 600 ºC
Voltage 0-50 mV...................................- 1,999 to 9,999 (Adjustable)
All input types are factory calibrated. NBR 12771, Pt100 NBR
13773/97 (IEC-751) (385) calibrated thermocouples.
Accuracy: Thermocouples:..............................0.25 % F.S. ± 1 ºC
Pt100:................................................0.2 % F.S. ± 1 ºC
0-50 mV: .........................................0.2 % F.S. ± 1
digit
Resolution:......................................................................................0.1 ºC
Sampling rate:.............................................................2 timer per second
Input impedance: ...............................> 10 M (Thermocouples, 50 mV)
Pt100 excitation current:............................................................0.165 mA
Power supply:......................................................................9 Vdc battery
AC / DC adapter .............................................................................6 Vdc
Estimated autonomy:.................................... 400 hours (alkaline battery)
Dimensions:................................................................. 165 x 73 x 36 mm
Weight:.............................................................................................205 g
Protection
: ..........................................................................................IP40
3 CONNECTIONS
Female input connectors suitable for RTD / TC compensated
connectors, as shown below.
Figure 01 — Meter supported connectors
Only two input terminals are used to connect thermocouples. Input
terminals have different size holes, which prevent inappropriate
installations.
Figure 02 — Thermocouple connection
All three terminals are used for Pt100 (RTD)
Figure 03 – Pt100 connectors sensors
The Pt100 internal connections should be made as shown in figure
below.
Figure 04 — Internal connections of Pt100 sensors
For the 0-50 mV signal, installation should comply with the following
connections:
Figure 05 — 0-50 signal connection
Zobrazit stránku 0
1 2 3 4

Shrnutí obsahu

Strany 1 - SMARTMETER

NOVUS PRODUTOS ELETRÔNICOS LTDA. – WEBSITE: WWW.NOVUSAUTOMATION.COM - TEL: +55-51-3323-3600 1/4 SMARTMETER PORTABLE TEMPERATURE METER - INSTRUCTION

Strany 2 - 4 OPERATION

SmartMeter NOVUS PRODUTOS ELETRÔNICOS LTDA. 2/4 4 OPERATION The meter features two input channels. Channel readouts are updated two times pe

Strany 3 - Table 1

SmartMeter NOVUS PRODUTOS ELETRÔNICOS LTDA. 3/4 CONFIGURATION CYCLE To enter the configuration cycle, press BACK and SET . The meter displa

Strany 4

SmartMeter NOVUS PRODUTOS ELETRÔNICOS LTDA. 4/4 #$$#$$#$$#$$(Low Calibration Channel 2) INPUT OFFSET CALIBRATION: Adjusts the lowest limi

Komentáře k této Příručce

Žádné komentáře